The machine says: Wide
QRS rhythm, left axis deviation. Do you have a different interpretation?
Figure 1
This is a paced-rhythm
(60 per minute). Pacer artifacts are best seen in v4 and V5. Bipolar pacing
will generate tiny artifacts unlike unipolar pacing. We can also see P waves in
V1 and V2 (intrinsic SR with CHB).
A frequently asked
question:
“I don’t see pacemaker
spikes”
1. The pacemaker may not
be pacing… usually due to inhibition by the patient’s intrinsic heart rate
(which may be faster than the lower limit of the pacemaker).
2. The pacemaker may be
pacing, BUT the stimulus artifacts (spikes) are not visible:
UNIPOLAR PACING STIMULUS
ARTIFACTS are large and nearly always visible. BIPOLAR PACING STIMULUS
ARTIFACTS may be too small in amplitude to see. In unipolar leads, the current
flows between the pulse generator and the tip of the lead (large area). So it
creates a large stimulus artifact. For bipolar pacing, the distance is only
between the tip and the ring which is only about a cm. So, the pacing stimulus
may be very difficult to see on surface ecg.
Reference:
Chow AC and Buxton AE.
2006. Defibrillators: All You Wanted to Know. Malden, MA, Blackwell Pub Inc
#385
No comments:
Post a Comment
Note: Only a member of this blog may post a comment.